INSTALLATION CONSIDERATIONS
TO ASSURE PROPER MAG METER READINGS.

In order to avoid errors due to air it is important that you avoid the installations shown at right.  The meter installed on the top of the pipe may contain

maginstall1.gif (18458 bytes)

Horizontal pipe run
Install in slightly ascending pipe section.  If not possible assure adequate velocity to prevent air, gas or vapor from collecting in the upper part of teh flow tube.

maginstall2.gif (9167 bytes)

Open Feed or discharge Install meter in low section of the pipe.

maginstall3.gif (6977 bytes)

If you have a down pipe over 16 feet in length install air valve downstream of flowmeter.

maginstall4.gif (11006 bytes)

Always install shut-off valves downstream of the flowmeter.

maginbstall5.gif (9141 bytes)

Never install mag meter on the suction side of a pump.

maginstall6.gif (5578 bytes)